Medical Device Recall: Battery Charging Station; Model: 0998-00-0802;
Datascope Corp. · February 6, 2026
Reason for Recall
The Battery Charging Station is an optional dual-bay charger used to charge the Cardiosave IABP Lithium- Ion batteries when they are not being used to operate the IABP. Getinge has identified that a protruding screw in the left battery bay limits full insertion of certain batteries and prevents proper charging. The right bay of the Battery Charging Station is not affected by this issue.
Distribution
Worldwide - US Nationwide distribution in the states of FL, DC, GA, IL, IN, MD, MI, NC, TX, WA, WI and the countries of Australia, Hong Kong, Israel, Japan, Norway.
States Affected
NATIONWIDE
Quantity Affected
28 units (20 US, 8 OUS)
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1587-2026
Status: ongoing