Class II

Medical Device Recall: Et Control, Software option for anesthesia gas machine to achieve and maintain targeted end tidal oxygen and anesthetic agents, Field replaceable unit (FRU) part numbers: 1011-8000-000-S, 5700665

Datex-Ohmeda, Inc. · July 30, 2026

Reason for Recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Distribution

Worldwide - US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

410 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3192-2026

Status: ongoing