Class I
Medical Device Recall: Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M1091607-R was installed during the last Preventative Maintenance.
Datex-Ohmeda Inc · May 8, 2026
Reason for Recall
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Distribution
Pending
Quantity Affected
550 (367 US; 183 OUS)
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2359-2026
Status: ongoing