Class I

Medical Device Recall: Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), STK-DO-013 (MT27409-1), STK-DO-103 (MT27409-2), STK-DO-109 (MT27409-2)

Dexcom, Inc. · May 12, 2025

Reason for Recall

Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead to untreated hypo or hyperglycemia which can cause seizures, vomiting, loss of consciousness, or death.

Distribution

worldwide distribution - US Nationwide and the countries of Andorra, Argentina Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Greece, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Latvia, Liechtenstein, Lithuania, Luxembourg, Malta, Netherlands, New Zealand, Norway, Oman, Poland, Romania, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.

States Affected

NATIONWIDE

Quantity Affected

38,127 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1946-2025

Status: ongoing