Class II

Medical Device Recall: Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones. Part Number:

EBI, LLC · March 22, 2024

Reason for Recall

Damage to the lead wire sheath (black cable) component and causing exposure of the wires, potentially lead to no therapeutic treatment signal being delivered to the electrodes, and result in delay of treatment

Product Description

Biomet OrthoPak Non-invasive Bone Growth Stimulator System-Indicated for the treatment of an established nonunion acquired secondary to trauma, excluding vertebrae and all flat bones. Part Number: 1067718

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

1412 units ( 2 leads per assembly)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1617-2024

Status: ongoing