Class II

Medical Device Recall: MOSAIQ Oncology Information System with Particle Therapy License (License Kit 45700000000IQRO)

Elekta, Inc. · December 10, 2025

Reason for Recall

Using oncology information system that manages workflow may result in overtreatment if:Plan delivered with Particle Therapy IHE-RO TDW-II interface,With one field per fraction or last field partial treatment,Partial treatment occurs followed by treatment field interruptions with manually recorded partial treatment,Then treatment unit doesn't check meterset value so partial treatment delivered twice

Distribution

US Nationwide distribution in the state of TX.

States Affected

NATIONWIDE

Quantity Affected

1

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1078-2026

Status: ongoing