Class II

Medical Device Recall: Brand Name: DJO SURGICAL Product Name: ALTIVATE REVERSE WEDGE GLENOID REAMER BOSS DRILL, 6.5mm Model/Catalog Number: 804-06-312 Product Description: Material: S.S., Non Sterile

Encore Medical, LP · July 18, 2025

Reason for Recall

Their is a potential that the reamer may kick or bind up during or immediately prior to use.

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CO, FL, HI, IA, IL, IN, KS, KY, LA, MD, ME, NE, NJ, NY, OH, PR, RI, SC, TN, TX, UT, VA, WA, WI.

States Affected

NATIONWIDE

Quantity Affected

907 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2547-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.