Class II

Medical Device Recall: SlideView VS200 slide viewer,

EVIDENT SCIENTIFIC INC · July 22, 2024

Reason for Recall

A broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made contact with the housing/frame and caused a short circuit and destroyed the fuse in the laser. This caused the shutter to remain in its most recent state; either in an open or closed position. Once the fuse was destroyed the shutter could not be operated.

Distribution

US

Quantity Affected

N/A

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2416-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.