Medical Device Recall: Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R;
Exactech, Inc. · April 26, 2024
Reason for Recall
The AcuMatch L-Series 22mm Inner Diameter Bipolar Hip Liner lots were packaged without the specified ethylene vinyl alcohol (EVOH) layer. Between 2004 and August 2021, our packaging process utilized two different types of packaging materials: 1) Low Density Polyethylene (LDPE), Nylon, and EVOH, or 2) LDPE and Nylon without EVOH.
Product Description
Acumatch L-Series BIPOLAR LINER, 22mm (BIPOLAR ENDOPROSTHESIS-Hip), Item Numbers: a) 100-22-19, SZ. L; b) 100-22-20, SZ. M; c) 100-22-21, SZ. N; d) 100-22-22, SZ. P; e) 100-22-23, SZ. R; f) 100-22-24, SZ. S; g) 100-22-25, SZ. T
Distribution
Worldwide - US distribution in Puerto Rico and the countries of Great Britain, Japan.
States Affected
IN
Quantity Affected
589 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2236-2024