Class II

Medical Device Recall: Exactech Equinoxe REVERSE SHOULDER,36mm Humeral Liner: a) +0mm, Item Number 320-36-00, b) +2.5mm, Item Number 320-36-03; Shoulder Arthroplasty

Exactech, Inc. · March 6, 2024

Reason for Recall

The packaging of these affected UHMWPE humeral liners and glenoids are nonconforming as they do not meet the established packaging specification. They were packaged in vacuum bags that did not contain an additional oxygen barrier layer consisting of Ethylene Vinyl Alcohol (EVOH).

Distribution

Nationwide and Australia, Austria, Belgium, Brazil, Chile, Canada , China, Colombia, France, Germany, Great Britain, Greece, Guatemala, India, Italy, Japan, Korea, Luxembourg, Malaysia, Netherlands, New Zealand, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand

States Affected

NATIONWIDE

Quantity Affected

11863 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1416-2024

Status: ongoing