Medical Device Recall: Airvo 2 Humidifier, REF: PT101US, and myAirvo 2 Humidifier, REF: PT100US
Fisher & Paykel Healthcare, Ltd. · April 2, 2024
Reason for Recall
Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
States Affected
NATIONWIDE
Quantity Affected
7,147
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1823-2024
Status: ongoing