Class II
Medical Device Recall: Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Fort Defiance Industries, LLC · July 8, 2026
Reason for Recall
The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
Distribution
Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
States Affected
NATIONWIDE
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2983-2026
Status: ongoing