Class II

Medical Device Recall: Optima MR360, NMRI system

GE Healthcare (China) Co., Ltd. · January 31, 2025

Reason for Recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Distribution

Domestic US distribution nationwide. International distribution worldwide.

States Affected

NATIONWIDE

Quantity Affected

33