Class II

Medical Device Recall: Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms

GE Healthcare Finland Oy · September 26, 2025

Reason for Recall

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Distribution

Worldwide - US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

5990

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0456-2026

Status: ongoing