Medical Device Recall: SIGNA Premier systems
GE Healthcare LLC · February 27, 2026
Reason for Recall
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
States Affected
NATIONWIDE
Quantity Affected
41
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1767-2026
Status: ongoing