Medical Device Recall: GE Healthcare MAC VU360, REF# part number 2030361-023; Electrocardiograph
GE Medical Systems Information Technologies Inc · June 29, 2026
Reason for Recall
Software issue issue with the MAC VU360 v3.0 system. After selecting the order refresh icon, an open preview or review screen might fail to close automatically after 2 minutes of inactivity. As a result, data from a previous patient may remain on the screen and could be incorrectly associated with a subsequent patient.
Distribution
Worldwide distribution - US Nationwide and the countries of Belgium, Hong Kong, Netherlands, Sweden, UK.
States Affected
NATIONWIDE
Quantity Affected
163 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2876-2026
Status: ongoing