Class II

Medical Device Recall: PORTRAIT CORE SERVICES SOFTWARE V1.1, Ref# 5700209

GE Medical Systems Information Technologies Inc · July 20, 2026

Reason for Recall

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.

Distribution

US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.

States Affected

MI, NY

Quantity Affected

23 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3036-2026

Status: ongoing