Class II

Medical Device Recall: Starcam/Camstar; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING · September 12, 2025

Reason for Recall

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Distribution

Worldwide

Quantity Affected

102 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0303-2026