Class II

Medical Device Recall: OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional too

GE OEC Medical Systems, Inc · November 21, 2024

Reason for Recall

Due to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.

Product Description

OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

Distribution

U.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden, Switzerland, and United Kingdom

States Affected

AK, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Quantity Affected

100 systems

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0704-2025

Status: ongoing