Class II

Medical Device Recall: Gentuity HF-OCT Imaging System Model G10-01; Software version 23.2.13;

Gentuity, LLC · March 23, 2026

Reason for Recall

The potential for repeated frames during HF-OCT pullback when using the Gentuity HF-OCT Console (software version 21.11 to 23.3.13) with the Vis-Rx and Vis-Rx Prime catheters. This correction is intended to ensure that users are fully informed of how to recognize, avoid, and mitigate this condition to maintain accurate longitudinal (length) measurements during OCT imaging.

Distribution

Domestic - US Nationwide distribution in the states of CA, FL, GA, IL, MA, MN, NJ, NY, OH, PA, TX; International: United Arab Emirates, Belgium, Switzerland, Cyprus, Denmark, Spain, Finland, France, Greece, Italy, Japan, Kuwait, Latvia, Poland, Saudi Arabia, Slovakia;

States Affected

NATIONWIDE

Quantity Affected

184 units (28 US, 156 OUS)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1823-2026

Status: ongoing