Medical Device Recall: Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Getinge Disinfection Ab · August 29, 2025
Reason for Recall
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Distribution
Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
States Affected
NATIONWIDE
Quantity Affected
902 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0016-2026
Status: ongoing