Medical Device Recall: Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.
Hamilton Medical AG · May 29, 2026
Reason for Recall
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
Distribution
US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.
States Affected
NATIONWIDE
Quantity Affected
27,270 pcs
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2463-2026
Status: ongoing