Medical Device Recall: Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and opt
Hamilton Medical AG · June 3, 2025
Reason for Recall
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.
Product Description
Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
1040 systems
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2328-2025
Status: ongoing