Class I

Medical Device Recall: Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and opt

Hamilton Medical AG · June 3, 2025

Reason for Recall

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

Product Description

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

1040 systems

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2328-2025

Status: ongoing