Class II

Medical Device Recall: HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-0

HeartSine Technologies Ltd · February 14, 2024

Reason for Recall

Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.

Product Description

HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10

Distribution

Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.

States Affected

NATIONWIDE

Quantity Affected

12

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1141-2024

Status: ongoing