Class II

Medical Device Recall: I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Strai

I.T.S. GmbH · January 23, 2026

Reason for Recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Product Description

I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.

Distribution

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

States Affected

NATIONWIDE

Quantity Affected

510 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1598-2026

Status: ongoing