Class II
Medical Device Recall: Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
I.T.S. GmbH · January 23, 2026
Reason for Recall
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
States Affected
NATIONWIDE
Quantity Affected
11 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1600-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.