Class II

Medical Device Recall: ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump

ICU Medical, Inc. · June 10, 2026

Reason for Recall

In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

States Affected

NATIONWIDE

Quantity Affected

12525 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2723-2026

Status: ongoing