Medical Device Recall: ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
ICU Medical, Inc. · December 15, 2025
Reason for Recall
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
States Affected
NATIONWIDE
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1120-2026
Status: ongoing