Class II

Medical Device Recall: LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

ICU Medical, Inc. · November 24, 2025

Reason for Recall

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

Distribution

US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

States Affected

NATIONWIDE

Quantity Affected

12 customers

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0966-2026

Status: ongoing