Medical Device Recall: LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
ICU Medical, Inc. · November 24, 2025
Reason for Recall
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Distribution
US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
States Affected
NATIONWIDE
Quantity Affected
12 customers
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0966-2026
Status: ongoing