Medical Device Recall: Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960
ICU Medical, Inc. · March 24, 2026
Reason for Recall
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
Distribution
US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX, FL, AR, CT, PA, WA, WI, CO, VT, NY, NH, KY, IA, KS, AL, NM, IL, MO, NJ.
States Affected
NATIONWIDE
Quantity Affected
20,506
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2210-2026
Status: ongoing