Medical Device Recall: Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount
IHB OPERATIONS B.V. · March 27, 2025
Reason for Recall
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, FL, IA, ID, IL, LA, MA, MD, MI, MN, NC, NH, NJ, NM, NY, PA, TN, TX, UT, VA, WI, WV.
States Affected
NATIONWIDE
Quantity Affected
112 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1663-2025
Status: ongoing