Class II

Medical Device Recall: Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Descri

Insightra Medical Inc · July 19, 2025

Reason for Recall

Device contains indications for use and device compatibility claims that have not be reviewed and approved for safety and effectiveness by the FDA. Instructions for use are not consistent with product training.

Product Description

Brand Name: Ultra IABP Catheter Kit Product Name: Ultra IABP Catheter Kit Model/Catalog Number: IMU7F-40, IMU7F-35, IMU7F-30, IMU7F-25, and IMU7F-20 Software Version: Not applicable Product Description: The Insightra Ultra IABP Catheter Kit is a 7 French catheter available with either a 40, 35, 30, 25, or 20 cc balloon. The catheter has two lumens and is designed to provide counterpulsation cardiac assist therapy. The outer lumen is a channel for helium used to inflate and deflate the balloon. The inner lumen is used for a guidewire and blood pressure measurement. It is packaged as a convenience kit with accessories (one-way valve, 50 cc syringe, guide wires, introducer needle, dilator, pressure tubing, Datascope & Arrow adaptors). The balloon inflation and deflation cycle is synchronized (using a commercially available control console) with the ECG or arterial pressure to provide counterpulsation synchronized with the heartbeat. It is intended to increase coronary perfusion, decrease the workload of the left ventricle and allow healing of the myocardium Component: No

Distribution

Worldwide - US Nationwide distribution in the states of TX, MO, FL, MI and the countries of Albania, Azerbaijan, Bangladesh, Brazil, Cambodia, Colombia, Hong Kong, India, Indonesia, Iran, Israel, Saudi Arabia, Kuwait, Lebanon, Malaysia, Nepal, Oman, Pakistan, Pakistan, Peru, Romania, Russia, Serbia, Singapore, United Arab Emirates.

States Affected

NATIONWIDE

Quantity Affected

24227 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2298-2025

Status: terminated