Class II

Medical Device Recall: da Vinci 5 Surgeon Console Viewer Display, ASSY, DV5 CONSOLE, IS5000, Part Number: 380730-45

Intuitive Surgical, Inc. · December 30, 2025

Reason for Recall

Robotic-assisted surgical system has Viewer Display manufactured with outdated firmware, may result in loss of one display-loss of 3D surgical field (depth perception) but 2D maintained, or loss of both displays-complete visualization loss, without fault, instruments remain active, could lead to instrument contact with tissue, resulting in injury/bleeding; may require alternate surgical modality.

Distribution

US Nationwide distribution in the states of KS, AZ, MA, FL, MI, NC, IL, DC, TX, KY, MS, CA, MN, NV, NY, WI, ID, GA, CT, SC, UT, VA.

States Affected

NATIONWIDE

Quantity Affected

47

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1355-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.