Class II

Medical Device Recall: Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495

Kreatech Biotechnology Bv · November 1, 2024

Reason for Recall

The DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.

Distribution

US States: FL, NY

States Affected

FL, NY

Quantity Affected

23 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0522-2025

Status: ongoing