Class II

Medical Device Recall: Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 8811

Laerdal Medical Corporation · January 13, 2026

Reason for Recall

units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

Product Description

Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;

Distribution

Worldwide distribution - US Nationwide and the countries of Denmark, Sweden, Norway, Canada, Finland, Australia, Spain, France, United Kingdom, Italy, Korea, Germany, Netherlands, Switzerland, New Zealand, Singapore, Malaysia, Austria.

States Affected

NATIONWIDE

Quantity Affected

1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1341-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.