Class II

Medical Device Recall: Leica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for freezing and sectioning tissue samples

LEICA BIOSYSTEMS NUSSLOCH GMBH · September 4, 2024

Reason for Recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

322 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0087-2025

Status: ongoing