Class II

Medical Device Recall: NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 4

Mc3 Inc · October 11, 2023

Reason for Recall

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

Product Description

NautilusTM Smart ECMO Module with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: 48135, 48135E

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

188 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1310-2024

Status: ongoing