Class II

Medical Device Recall: Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER S

Medline Industries, LP · January 8, 2026

Reason for Recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Product Description

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

14,525

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1465-2026

Status: ongoing