Class II

Medical Device Recall: Medline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7315 NEURO PACK-LF DYN

Medline Industries, LP · April 27, 2026

Reason for Recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Product Description

Medline convenience kit: BHS NEURO DYNJ904900N MEDTRX NEURO PACK DYNJ67267B MRI DBS PACK DYNJ9814630O NEURO CDS-LF CDS983383L NEURO PACK DYNJ59397Q DYNJT7315 NEURO PACK-LF DYNJ25097I NEURO VP SHUNT/VAGAL NERVE-RF DYNJ21925S PK NEURO SHUNT BUMC DYNJT6428 RR-NEURO DBS HEAD PACK-LF DYNJ0785736M VP SHUNT SYNJ10307B

Distribution

US Nationwide distribution. OUS distribution pending.

States Affected

NATIONWIDE

Quantity Affected

49,654 kits total

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2428-2026

Status: ongoing