Medical Device Recall: Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators LABOR CDS CDS981762B VAG DELIVERY ADMISSION BAG DYKA1054F
Medline Industries, LP · July 20, 2026
Reason for Recall
Convenience kits contain BD ChloraPrep Clear - 1mL Applicators from specific lots that were recalled by the manufacturer because they may have an open or incomplete seal on the applicator packaging. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
222,653 total
Source: FDA (openFDA Device Enforcement)
Recall number: Z-3076-2026