Class II

Medical Device Recall: Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530A

Medline Industries, LP · May 28, 2026

Reason for Recall

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Distribution

US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.

States Affected

NATIONWIDE

Quantity Affected

9,272 kits

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2612-2026

Status: ongoing