Class II

Medical Device Recall: Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit. Respiratory gas humidifier.

Medline Industries, LP · July 13, 2026

Reason for Recall

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Distribution

Worldwide distribution - US Nationwide and the country of Panama.

States Affected

NATIONWIDE

Quantity Affected

20 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2991-2026

Status: ongoing