Medical Device Recall: Medline Kits containing Tego Connectors: Medline IR Pack, SKU DYNJ67205C Medline NO CATHETER DIALYSIS INSERTION, SKU CVI5200

Medline Industries, LP · January 8, 2026

Reason for Recall

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

516

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1463-2026

Status: ongoing