Class II

Medical Device Recall: Medline medical procedure convenience kits labeled as: 1) THORACIC ROBOTICS, REF DYNJ908777B; 2) CARDIAC ROBOTIC WHC KIT-LF, REF DYNJ911359

Medline Industries, LP · December 16, 2025

Reason for Recall

Medline kits containing Stryker's Strykeflow 2 Suction Irrigators due to a change in design in 2024, irrigation solution is able to travel to the handpiece and battery pack during use of the Strykeflow 2 Suction Irrigator. This design change has caused some devices to leak and/or emit vaporized saline (which can resemble smoke) during use.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

96 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1085-2026

Status: ongoing