Medical Device Recall: MEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR PACK, etc. (see the
Medline Industries, LP · May 4, 2026
Reason for Recall
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Product Description
MEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR PACK, etc. (see the recall documents for a full list of affected products)
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
States Affected
NATIONWIDE
Quantity Affected
44656 units