Class II

Medical Device Recall: MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit

Medline Industries, LP · May 4, 2026

Reason for Recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Product Description

MEDLINE Medical Procedure Kits labeled as: 1) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775I; 2) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775J; 3) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775K; 4) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775L; 5) RETINA LAC/USC CDS, Medline Kit Number/SKU CDS983775M; 6) TONSIL PACK, Medline Kit Number/SKU DYNJ84480; 7) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080; 8) VITRECTOMY, Medline Kit Number/SKU DYNJVB1080A.

Distribution

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

States Affected

NATIONWIDE

Quantity Affected

2340 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2507-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.