Class II

Medical Device Recall: MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515

Medline Industries, LP · May 4, 2026

Reason for Recall

During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.

Product Description

MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DYND13614 - 14 Fr; 11. DYND13616 - 16 Fr.

Distribution

Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.

States Affected

NATIONWIDE

Quantity Affected

2237471 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2479-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Medline Industries, LP Medical Device Recall: MEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 | SafeCheck