Class I

Medical Device Recall: (1) Nursing Skills kit, SKU #EDUC05038B, Case UPC 40195327019878, Non-sterile, For Training Purposes Only, Not For Human Use.

MEDLINE INDUSTRIES, LP - Northfield · October 16, 2023

Reason for Recall

The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.

Distribution

US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.

States Affected

NATIONWIDE

Quantity Affected

172 kits

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0562-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.