Class II

Medical Device Recall: Medline convenience kits containing semi-rigid suction liners labeled as follows: DENTAL, Pack Number DYNJ909451A

MEDLINE INDUSTRIES, LP - Northfield · June 27, 2024

Reason for Recall

Due to reported adverse events, Medline Industries, LP. is issuing a recall on specific SKUs and lots of 1500 cc semi-rigid liners manufactured from 09/2023-11/2023 due to the potential loss of suction or low suction during intermittent suction use.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Panama.

States Affected

NATIONWIDE

Quantity Affected

19 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3108-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.