Class I

Medical Device Recall: Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5 mm ID; (3) REF DYNJ18870, Polyurethane Cuff 7.

MEDLINE INDUSTRIES, LP - Northfield · February 26, 2024

Reason for Recall

Complaints have been received that the device inflation tube detached and/or tore from the main tube, resulting in potential moisture buildup, loss of pressure, or inability to inflate. There were also reports the suction pump is difficult to connect or detaches during use.

Product Description

Medline Sub-G Endotracheal Tube with Subglottic Suctioning: (1) REF DYNJ18860, Polyurethane Cuff 6.0 mm ID; (2) REF DYNJ18865, Polyurethane Cuff 6.5 mm ID; (3) REF DYNJ18870, Polyurethane Cuff 7.0 mm ID; (4) REF DYNJ18875, Polyurethane Cuff 7.5 mm ID; (5) REF DYNJ18880, Polyurethane Cuff 8.0 mm ID; and (6) REF DYNJ18885, Polyurethane Cuff 8.5 mm ID. Medline SubG Endotracheal Tube with Subglottic Suctioning packaged into the following convenience kits: (1) Medline Intubation Kit - Drawer 3, Reorder No. ACC010502, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (2) Medline Intubation Tube - Drawer 3, Reorder No. ACC010527, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 6 kits/case; (3) Medline Adult Intubation, Reorder No. ACC010540, containing 8.5 size subglottic cuff, 3 kits/case; (4) Medline Adult Intubation, Reorder No. ACC010540A, containing 8.5 size subglottic cuff, 3 kits/case; (5) Medline Airway Kit - Drawer 3, Reorder No. ACC010717, containing 6.0, 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 3 kits/case; (6) Medline Airway Kit, Reorder No. DYNDJ1132, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 26 kits/case; (7) Medline Adult Intubation W Evac, Reorder No. DYNDJ1133, containing 7.0, 7.5, and 8.0 sizes subglottic cuff, 10 kits/case; (8) Medline Backup Kit, Reorder No. DYNJ909501, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (9) Medline Backup Kit, Reorder No. DYNJ909501A, containing 6.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 1 kit/package; (10) Medline Adult Intubation Tray, Reorder No. DYNJAA269, containing 7.0, 7.5, 8.0, and 8.5 sizes subglottic cuff, 10 kits/case.

Distribution

Distribution was nationwide. There was government/military distributon. Foreign distribution was made to San Salvador and United Arab Emirates.

States Affected

NATIONWIDE

Quantity Affected

168,632 tubes and 13,092 kits

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1876-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.